An Air-Blank test is performed when you want to confirm that the SCRAM Remote Breath Pro device’s fuel cell has not been damaged or contaminated by an outside source. The Air-Blank test results should reflect a zero BrAC level.
Initiate the Air-Blank Test in SCRAMNET
- Select the Inventory button.
- Enter the serial number of the device in the Serial Number field and select the Search button.
- Select the Serial Number hyperlink.
- On the Device Details page, select the Initiate Air-Blank button.
- Power on (In Inventory) or wake up (Assigned to Client) the device. No further action is required on your part, the device will display the following:
-
- PLEASE WAIT
Once the air-blank test is complete, the device will return to the pre-test state: “Device Unassigned” for inventory devices, or “Monitoring” for assigned devices.
- Verify that the air-blank test results appear on the Device Details page in SCRAMNET. On the Device Details page, select the Refresh Page to Latest Status button. A record of the Air-Blank test appears in the Calibration History section.